Try searching by key words, order- or product code or serial number e.g. “CM442” or “Technical information”
Enter at least 2 characters to start the search.
Condumax CLS16D is a digital conductivity probe for hygienic applications in pure & ultrapure water

Digital conductivity sensor
Condumax CLS16D

Memosens conductive conductivity sensor for hygienic applications in pure and ultrapure water

CLS16D
Please Wait (spinning wheel)
Loading price
Prices temporarily unavailable
Price on request
activating price
price: provide information
from CHF ??.-
Filter Measuring / Application task
Measuring / Application task

Configure to download the CAD drawings of this product

  • Measuring principle

    Conductive

  • Application

    Pure and ultrapure water, pharmaceutical industry

  • Characteristic

    Hygienic 2-electrode conductivity sensor

  • Measurement range

    k=0,1: 0,04 to 500 µS/cm

  • Measuring principle

    Conductivity cell

  • Design

    Hygienic 2-electrode conductivity sensor with coaxially arranged electrodes, electropolished stainless steel and FDA sealing material; sterilizable

  • Material

    Isolation: PEEK
    Electrode: polished stainless steel 1.4435
    Sealing: Isolast (FFKM) FDA conform

  • Dimension

    Electrode diameter: 17 mm (0.66 inch)
    Electrode length: 54 mm (2.13 inch)

  • Process temperature

    -5 to 120 °C (23 to 248 °F)
    For Sterilizsation: max. 150 °C at 5 bar
    (Max. 302 °F at 73 psi)

  • Process pressure

    13 bar at 20 °C (188 psi at 68 °F)
    9 bar at 120 °C (130 psi at 248 °F)

  • Temperature sensor

    NTC 30K

  • Connection

    Process connection: Varivent, Neumo, Clamp
    Cable: Memosens connector

  • Ingress protection

    IP68

  • Additional certifications

    Material certification 3.1
    EHEDG certified, hygienic design, CoC ASME BPE
    Conformity to cGMP derived requirements
    ASME BPE
    FDA approved
    Regulation (EC) No.
    1935/2004
    China Food