This product is no longer available.
Please check new generation in the section predecessor / successor or contact your local Endress+Hauser entity.
-
Successor
-
Standard products
- Reliable, robust and low-maintenance
Technical excellence
Simplicity
-
Specialized products
- Designed for demanding applications
Technical excellence
Simplicity
Variable
FLEX selections Technical excellence Simplicity - F
- L
- E
- X
Fundamental selection
Meet your basic measurement needs
Technical excellence Simplicity - F
- L
- E
- X
Lean selection
Handle your core processes easily
Technical excellence Simplicity - F
- L
- E
- X
Extended selection
Optimize your processes with innovative technologies
Technical excellence Simplicity - F
- L
- E
- X
Xpert selection
Master your most challenging applications
Technical excellence Simplicity Variable
-
-
Measuring principle
Thermocouple
-
Characteristic / Application
S.A. Style
universal range of application
usable with or without a thermowell
spring loaded for use in thermowell
Options available to construct removable inserts (spare parts) -
Thermowell / protection tube
to use with or without thermowell (can be ordered separately)
-
Insert / probe
mineral insulated (MI), flexible
-
Max. immersion length on request
up to 2.000 mm (78,74'')
-
Material protection tube/ thermowell
1.4401 (316)
1.4841 (310)
Alloy 600 (2.4816) -
Optional coating
Not defined
-
Process connection
Union nipple:
NPT 1/2"
G1/2" -
Tip shape
straight
-
Operating temperature range
Type J:
-210 ...760 °C
(-346 ...1.400 °F)
Type K:
-270 ...1.100 °C
(-454 ...2.012 °F)
Type N:
-270 ...1.100 °C
(-454 ...2.012 °F)
Type T:
-270 ...400 °C
(-454 ...725 °F) -
Max. process pressure (static)
at 20 °C: 10 bar (150 psi) without thermowell
-
Accuracy
class 1 acc. to IEC 60584
-
Response time
depending on configuration
-
Integration head transmitter
yes (4 ... 20 mA; HART; PROFIBUS PA; FOUNDATION FIELDBUS)
-
Certification
EN10204 3.1 Material traceability